SUNRISE MEDICAL PRIVADA MISIONES ZIPPIE VOYAGE; STROLLER (CHAIR WITH CASTERS) Back to Search Results
Model Number EIZ14
Device Problem Detachment of Device or Device Component (2907)
Patient Problems Swelling (2091); Tooth Fracture (2428)
Event Date 12/18/2017
Event Type  Injury  
Manufacturer Narrative
Sunrise medical has requested with the dealer that the stroller be returned for a full evaluation.The stroller has the slide-n-lock base in which the seating system will slide onto the base and lock into place.The seating system is successfully locked in place when an audible click is heard.Past evaluations have found that if there is an obstruction, the slide-n-lock base and the seating system will not lock and the locking click will not be heard.A replacement stroller was shipped to the dealer on (b)(6) 2017.If there are any relevant findings from the evaluation, a supplemental report will be filed.
 
Event Description
Per jp, the dealer, on (b)(6) 2017 while the child was at school the teacher was pushing the child around the corner and the seat slid off from the base with the child, hitting the ground.The child was taken to the hospital.The child allegedly has a chipped tooth and a "goose egg" (bump) on her forehead.
 
Manufacturer Narrative
The stroller in question was received and an evaluation was performed by a sunrise medical quality engineer on 02/20/2018.The following findings were reported from the evaluation: the stroller is a voyage vip model with a serial no.(b)(4) when unpacking the chair was observed that the support of the seating system was loose and also the screws of the base were loose.It was also noted that the harness straps are very long.For purposes of the test, the hitch movement was made in a similar way to how the user would mount the seating system and the straps were not caught.A test of the seating system was made attaching it to the base and adding 40 pounds of weight to simulate the weight of the user.Then a push test was done with the stroller on flat terrain as well as on a dirt road and confirmed that the seating system does not disengage from the base.The seating system did not detach from the base while testing on flat terrain as well on a dirt road.The alleged failure was unable to be duplicated.
 
Event Description
Please see initial mdr 9616084-2018-00001.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIPPIE VOYAGE
Type of Device
STROLLER (CHAIR WITH CASTERS)
Manufacturer (Section D)
SUNRISE MEDICAL PRIVADA MISIONES
no. 110 parque industrial
misiones de las californias
tijuana, baja california 22425
MX  22425
Manufacturer Contact
gustavo zambrano
2842 business park ave.
fresno, CA 93727
5592942840
MDR Report Key7197758
MDR Text Key97474596
Report Number9616084-2018-00001
Device Sequence Number1
Product Code INM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date 03/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberEIZ14
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/05/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age5 YR
Patient Weight14
-
-