The hill-rom technician found the power control board needed to be replaced.Per the hill-rom service manual the totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Preventative maintenance will minimize downtime due to excessive wear.Check for current leakage at the nurse call system communication connections.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The hill-rom technician replaced the power control board to resolve the issue.Based on this information, no further action is required.
|