The customer stated that they received erroneous results when testing one patient with two accu-chek inform ii meters (serial numbers(b)(4)).At 6:00 a.M., a sample was collected from the patient and tested with a laboratory method, resulting as 154 mg/dl.At 7:44 a.M., a sample from the patient was tested on meter serial number (b)(4), resulting as 511 mg/dl.At 7:46 a.M., a sample from the patient was tested on accu-chek inform ii meter serial number (b)(4), resulting as 201 mg/dl.At 7:46 a.M., a sample from the patient was tested on meter serial number (b)(4), resulting as 145 mg/dl.The customer also noted that they received a system error on this meter on (b)(6) 2018.The customer believed the 201 mg/dl value from meter serial number (b)(4) to be correct.No adverse events were alleged to have occurred with the patient.The patient was not seriously injured.Quality controls were tested on the meters within 24 hours and these passed.The customer's product was requested for investigation and replacement product was sent to the customer.Routine retention testing is performed.Retention testing data is reviewed and appropriate actions are taken as needed.
|