It was reported that during a splenic artery embolization, the coil did not properly shape.During removal, the coil unexpectedly stretched and detached.Both segments of the coil were removed together with the microcatheter from the patient.There was no reported intervention, health damage, or patient injury.
|
The device was returned for analysis.There was evidence of stretching along the proximal 15cm of the coil, with minimal damage in the remaining distal portion.There was no evidence of lost shape in the area that is not stretched or damaged.The implant coil portion that was returned was damaged and stretched; however, no fractures or breaks were noted along the length.The monofilament knot-tie was intact at the attachment zone, with evidence of stretching and prolapsing of the wire filar just distal to knot-tie.Based on the investigation, the implant coil was not broken into separate pieces; however, the coil was stretched and damaged.The root cause is unknown; however, the damage to the device exhibited evidence that it was subjected to excessive tensile force that exceeded the maximum strength specification.
|