Our evaluation findings for the returned 33300 outer tube: the shaft is bent, the insulation is cut and instrument failed high pot test, the distal tip is bent out of round, the locking channels in the distal tip that lock in the insert are broken off and were not returned.Because the customer said that one of the locking pieces fell into the patient and was removed, and the other one was not found to be in the patient, we suspect that it was most likely already missing before the outer tube was used in this procedure.The outer tube was manufactured in september 2007, so it is approximately 10 years old.Damages due to handling / wear and tear.
|
Allegedly, during a laparoscopic myomectomy procedure, the instrument outer tube broke inside the patient.Per the customer, "the clip was noted to be in the abdomen; the piece was retrieved, but second clip was not found.X-ray was done at the end of the case, which did not show retained item." hospital reports there was no negative patient impact.
|