The report of a severed pump cable was confirmed.The pump was returned assembled with the pump cable cut approximately 16 inches from the terminus of the pump end bend relief.It was reported that the surgeon mistakenly severed the pump cable during the patient's pump exchange.The distal end of the pump cable, measuring approximately 10 inches in length, was also returned.The modular cable was not returned.The sealed outflow graft and the sealed outflow graft bend relief were not returned.The apical cuff was not returned.Upon disassembly of the returned pump, examination of the pump blood-contacting surfaces revealed no depositions.The returned pump was cleaned, rebuilt, and functionally tested on a mock circulatory loop.The device operated as intended and in accordance with manufacturing specifications.A review of the device history records revealed no deviations from manufacturing or quality assurance specifications.The manufacturer is closing the file on this event.
|