• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Kinked (1339); Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/11/2018
Event Type  malfunction  
Event Description
It was reported that during implant procedure, the left ventricular lead had dislodged.The physician mentioned that prior to attempting to implant the lead, a kink was noticed.The lead was not used and another was implanted.The procedure was successful.The patient was stable.
 
Manufacturer Narrative
A complete lead was returned in one piece measuring 86.2cm.Analysis was normal.No anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7213884
MDR Text Key98046660
Report Number2017865-2018-01280
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Model Number1458Q/86
Device Lot NumberA000042921
Other Device ID Number05414734503198
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age64 YR
-
-