Catalog Number 295025-001
Device Problems Premature Discharge of Battery (1057); Battery Problem (2885)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 01/15/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because patients are provided with several onboard batteries, and the freedom driver has a redundant power source of external wall power.The freedom onboard battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia authorized distributor, reported that the freedom onboard battery discharged too quickly.The customer also reported that the patient was provided a replacement onboard battery.There was no reported adverse patient impact.
 
Manufacturer Narrative
Initial investigation and evaluation of the battery's (system management) smbus data revealed the battery to exhibit a low smbus full charge capacity and that it had accumulated a high cycle count.The battery was successfully recovered by use of the built-in zvchg recover utility, and then the battery was tested in accordance with the current evaluation process, where it failed the discharge test, thereby, confirming it did exhibit low capacity and confirming the customer-reported issue.The root cause of the reported "discharges too quickly" was the low capacity of the battery, likely because of the high cyclic usage the battery accumulated in use.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM ONBOARD BATTERY
Type of Device
BATTERY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7214964
MDR Text Key98419899
Report Number3003761017-2018-00039
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Remedial Action Replace
Type of Report Initial,Followup
Report Date 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number295025-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2018
Was the Report Sent to FDA? No
Date Manufacturer Received01/15/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1