Model Number 1153-
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/30/2017
Event Type  malfunction  
Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during the ventricular assist device (vad) implant procedure it was noted that the screw was missing from the strain relief outflow graft.The pump was not used and a different pump was implanted.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: hvad implant kit (b)(4), including the outflow graft, was returned for evaluation.A review of the manufacturing records confirmed that the implant kit met all requirements prior to release.The reported event was confirmed via visual inspection which revealed that, as received, a screw was missing from the outflow graft strain relief.The screw was not located within the returned implant kit packaging.Based on the available information, a possible root cause of the reported event may be attributed but not limited to an assembly error and/or improper handling of the device.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM-OUTFLOW GRAFT STRAIN RELIEF
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7215363
MDR Text Key98297720
Report Number3007042319-2018-00366
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00888707002639
UDI-Public00888707002639
Combination Product (y/n)N
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/31/2019
Device Model Number1153-
Device Catalogue Number1153-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2018
Date Manufacturer Received10/18/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age59 YR
Patient Weight106
-
-