Catalog Number 5100015250
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 09/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).One event was reported for this quarter.One device was received.One device evaluation is in progress.There were no remedial actions taken.This device is not labeled for single-use.
 
Event Description
This report summarizes 1 malfunction event in which an accessory broke in the device.There was no patient involvement; no patient impact.
 
Manufacturer Narrative
This mdr report is part of the malfunction summary reporting program, exemption number (b)(4).Supplemental rationale: 1 previously reported event is included in this follow-up record.Product return status: 1 device was received for evaluation.Evaluation status: 1 event was confirmed during testing.- 1 device was found to be affected by a broken collet.Additional information: 1 device is not labeled for single-use.1 device was not reprocessed and reused.
 
Event Description
This report summarizes 1 malfunction event in which an accessory broke in the device.There was no patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRODRILL SERIES STRAIGHT ATTACHMENT
Type of Device
UNIT, OPERATIVE DENTAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
zach baker
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key7219286
MDR Text Key98434617
Report Number0001811755-2018-00085
Device Sequence Number1
Product Code EIA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number5100015250
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/02/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-