Catalog Number 7209009000
Device Problems Device Remains Activated (1525); Sticking (1597); Positioning Problem (3009)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Date 09/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).One event was reported for this quarter.One device was received for evaluation; one event was confirmed.One device was found to be affected by corrosion in the switch and in the trigger assembly.There were no remedial actions taken.This device is not labeled for single-use.
 
Event Description
This report summarizes 1 malfunction event in which the device safety switch was stuck in the "run" position and the device had a sticky trigger.One reported event had no patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYSTEM 7 HIGH SPEED PRECISION SAW
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
zachary baker
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key7220184
MDR Text Key98526402
Report Number0001811755-2018-00313
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number7209009000
Was Device Available for Evaluation? No
Date Manufacturer Received09/02/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-