• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 SIG FEM ADPT TORQUE WRENCH; KNEE INSTRUMENT : TORQUE DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 SIG FEM ADPT TORQUE WRENCH; KNEE INSTRUMENT : TORQUE DEVICES Back to Search Results
Catalog Number 961673
Device Problems Disassembly (1168); Naturally Worn (2988)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/10/2018
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Revision total knee replacement dr (b)(6); (b)(6) hospital (b)(6) 2018.During use on patient, femoral torque wrench lodged on adaptor prosthesis, wrench very visually worn.Wrench struck with mallet and osteotomes to dislodge it in order to manage problem during procedure.10 mins delay to procedure, ae to patient, female patient initials (b)(6) date of birth (b)(6).
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # pc-(b)(4).Investigation summary the reported device was not returned to the investigation site for evaluation.The reported device has been evaluated at the australia facility and the reported event was confirmed.The black end cap was also noted to be missing.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIG FEM ADPT TORQUE WRENCH
Type of Device
KNEE INSTRUMENT : TORQUE DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46582-0988
5743725905
MDR Report Key7221040
MDR Text Key98328414
Report Number1818910-2018-52081
Device Sequence Number1
Product Code HXC
UDI-Device Identifier10603295239789
UDI-Public10603295239789
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number961673
Device Lot NumberSO2012915
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/05/2018
Date Device Manufactured04/04/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age81 YR
-
-