MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problems Unintended Arm Motion (1033); Mechanical Problem (1384); Unintended Movement (3026)
Patient Problem No Patient Involvement (2645)
Event Date 01/08/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.No parts have been received by the manufacturer for evaluation.
 
Event Description
A site representative reported that, while outside of a procedure the vertek arm would not lock in the proper way on the navigation system.There was no patient present when this issue was identified.No additional information was provided.
 
Manufacturer Narrative
The vertek arm was returned to the manufacturer for analysis.Analysis found that the arm remained loose on the starburst end when tightened by the user.Analysis found that the reported event was related to a mechanical issue.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7225739
MDR Text Key98487664
Report Number1723170-2018-00443
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450968
UDI-Public00613994450968
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 04/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733858
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-