• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Catalog Number BI70000028230
Device Problems Mechanical Problem (1384); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 01/08/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.It was reported that slides underneath the covers of the gantry were damaged.It was noted that the damage did not affect imaging or other functionalities of the imaging system other than x-direction movement.The bellows cover slides were then replaced to resolve the reported issue.The imaging system then passed the system checkout and was found to be fully functional.No parts have been received by the manufacturer for evaluation.
 
Event Description
A site representative reported that, while outside of a procedure, the gantry of the imaging system would intermittently not move in the x direction.There was no patient present when this issue was identified.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM 3RD EDITION
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7229524
MDR Text Key98635624
Report Number1723170-2018-00429
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K092564
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Health Professional
Type of Report Initial
Report Date 01/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70000028230
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/08/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/25/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-