• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913); Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/15/2018
Event Type  malfunction  
Manufacturer Narrative
The customer's reported complaint of a damaged head restraint was confirmed during visual inspection and was replaced.However, the reported complaint of a stopped compression was not confirmed or reproduced during these evaluation.The platform passed functional testing with no faults found.Review of the retrieved archive data found no significant discrepancies related to the reported stop compression complaint.Unrelated to the complaint, to ensure the autopulse platform is functional without issue, the bent battery lock clip, damaged enclosures, and missing screws were replaced.The autopulse platform is a reusable device and was manufactured in 2008 and has exceeded its expected service life of 5 years.Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for autopulse with serial (b)(4).
 
Event Description
During training the autopulse platform (sn: (b)(4)), was placed on continuous mode.The customer, however, reported the platform stopped compression multiple times for an unspecified reason.Additionally, the platform's head restraint cable was found damaged.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key7232829
MDR Text Key98864663
Report Number3010617000-2018-00122
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000017
UDI-Public00849111000017
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 01/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/15/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-