RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Material Discolored (1170)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Event Description
|
The customer reported that the ventilator screen was black and seeking part number for backlight inverter, battery and preventive maintenance kit.It was reported that the event occurred during clinical use.There was no report of patient harm or injury as a result of the event.
|
|
Manufacturer Narrative
|
Updated.The customer reported that the ventilator has been repaired.The backlight inverter resolved the issue.The determination could not be made that the device failed to meet specifications.The device was not being used for treatment when the reported event occurred, and there is not a relationship of the device to the reported problem.No parts returned to failure investigation; therefore, the root cause of the reported issue could not be determined.If parts are returned, the complaint will be reopened and an investigation will be performed.
|
|
Event Description
|
The customer reported that the ventilator screen was black and seeking part number for backlight inverter, battery and preventive maintenance kit.It was reported that the event occurred during clinical use.There was no patient involvement at the time of the event occurred.
|
|
Search Alerts/Recalls
|
|
|