The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device was returned to the manufacturer for analysis.The pusher was noted to have a bend in the hypotube section.The transition joint (hypotube-body coil) was broken at the laser weld into two halves; however, the pet was still intact.The implant was provided separately and noted to be in good condition, without any damage or deformities.The coil also had the attachment tether unsevered.The heater coil section was noted to be in conformance as well.The attachment tether demonstrated a stretched tail configuration, indicating non-thermal detachment.Based on the reported complaint and evaluation of the returned product, the stretched and broken coil indicated that excessive force had been applied to the device, which exceeded its maximum tensile specification.
|
It was reported that after several coils had been placed in a 13mm aneurysm, difficult advancement of the coil was encountered.During re-positioning and retraction of the coil, the coil detached in the microcatheter.Both segments of the coil were removed together with the microcatheter.There was no reported intervention or patient injury.The patient is reported to be doing well.
|