TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
|
Back to Search Results |
|
Model Number 4415 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/09/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).As per the user facility, the unit was turned off and on again resolving the problem.The unit will not be returned to the manufacturer as the problem was resolved.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the heater cooler would not make ice.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
(b)(4).The reported complaint was confirmed.The field service representative (fsr) observed the compressor switch to be turned on but the compressor was not running.The unit was restarted, the compressor came on but overheated and then shut off.Replacement parts are no longer available and the unit has reached its end of service life.The user facility has been advised to remove the unit from clinical use.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|