MAUDE Adverse Event Report: AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER Back to Search Results
Catalog Number SGC0301
Device Problems Break (1069); Difficult to Insert (1316); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/09/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The customer reported the device was discarded.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event Description
This event is filed for a cable break.It was reported that the patient, with degenerative mitral regurgitation (mr), underwent a mitraclip procedure.The steerable guide catheter (sgc) was prepared for use and no issues were noted.The sgc minus knob was turned a half turn to straighten the device; however, difficulty was noted during insertion into the patient anatomy.An additional half turn was made, 1 full turn.A noise was heard and the minus cable broke.The device was not used and a new sgc was prepared.The second sgc was inserted into the patient anatomy without difficulty and two clips were implanted, reducing the mr from grade 3+ to grade 1.There was no additional information provided.
 
Manufacturer Narrative
(b)(4).The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no other incidents reported from this lot.All available information was investigated and a definitive cause for the reported difficult to insert and cable break could not be determined.The reported device operating differently than expected (noise) appears to be related to the cable break.Based on the information reviewed, there is no indication of a product quality issue with respect to design, manufacturing or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
STEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
MENLO PARK, CA REG# 3005070406
abbott vascular
26531 ynez rd.
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key7236123
MDR Text Key98999855
Report Number2024168-2018-00768
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/08/2018
Device Catalogue NumberSGC0301
Device Lot Number70508U149
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age81 YR
Patient Weight53
-
-