• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1500400-23
Device Problems Difficult to Remove (1528); Improper or Incorrect Procedure or Method (2017); Folded (2630)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/12/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4) excessive force.The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that post stent implantation, negative was pulled on the 4.0x23 sierra stent delivery system (sds) for 10 seconds and then set on neutral pressure for the pullout.The balloon was noted to be folded flat.Strong resistance was met during retraction of the balloon through the guiding catheter and force was applied.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device was returned for analysis.The reported flat balloon fold was confirmed.The reported difficulty to remove was unable to be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation was unable to determine a conclusive cause for the flat balloon fold; however, the reported difficulty to remove appears to be related to circumstances of the procedure as it is likely the device interacted with the guiding catheter during retraction due to the flat balloon fold.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Manufacturer Narrative
(b)(4).Correction: the xience sierra is currently not commercially available in the u.S.; however, it is similar to a device sold in the u.S.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7236358
MDR Text Key98998284
Report Number2024168-2018-00781
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 04/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/09/2020
Device Catalogue Number1500400-23
Device Lot Number7100241
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/30/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-