Catalog Number 1500400-23 |
Device Problems
Difficult to Remove (1528); Improper or Incorrect Procedure or Method (2017); Folded (2630)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/12/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4) excessive force.The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that post stent implantation, negative was pulled on the 4.0x23 sierra stent delivery system (sds) for 10 seconds and then set on neutral pressure for the pullout.The balloon was noted to be folded flat.Strong resistance was met during retraction of the balloon through the guiding catheter and force was applied.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
|
Manufacturer Narrative
|
(b)(4).The device was returned for analysis.The reported flat balloon fold was confirmed.The reported difficulty to remove was unable to be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation was unable to determine a conclusive cause for the flat balloon fold; however, the reported difficulty to remove appears to be related to circumstances of the procedure as it is likely the device interacted with the guiding catheter during retraction due to the flat balloon fold.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|
Manufacturer Narrative
|
(b)(4).Correction: the xience sierra is currently not commercially available in the u.S.; however, it is similar to a device sold in the u.S.
|
|