STRYKER MEDICAL-KALAMAZOO LD304 MATERNITY BED; TABLE, OBSTETRICAL, AC-POWERED (AND ACCESSORIES)
|
Back to Search Results |
|
Catalog Number 4701000000 |
Device Problems
Mechanical Problem (1384); Unintended System Motion (1430)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/11/2018 |
Event Type
malfunction
|
Event Description
|
It was reported that while a patient was sitting on the labor and delivery bed, the bolt that attaches the foot crank weldment on the patient left side came out, causing the bed to unexpectedly lower.No adverse consequence or clinically relevant delay in treatment was reported.
|
|
Manufacturer Narrative
|
In addition to the missing bolt, the technician found the bushing was worn, affecting the base pivot spacer, bushing, cap screw, and bearing.
|
|
Event Description
|
It was reported that while a patient was sitting on the labor and delivery bed, the bolt that attaches the foot crank weldment on the patient left side came out, causing the bed to unexpectedly lower.No adverse consequence or clinically relevant delay in treatment was reported.
|
|