• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO LD304 MATERNITY BED; TABLE, OBSTETRICAL, AC-POWERED (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO LD304 MATERNITY BED; TABLE, OBSTETRICAL, AC-POWERED (AND ACCESSORIES) Back to Search Results
Catalog Number 4701000000
Device Problems Mechanical Problem (1384); Unintended System Motion (1430)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/11/2018
Event Type  malfunction  
Event Description
It was reported that while a patient was sitting on the labor and delivery bed, the bolt that attaches the foot crank weldment on the patient left side came out, causing the bed to unexpectedly lower.No adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
In addition to the missing bolt, the technician found the bushing was worn, affecting the base pivot spacer, bushing, cap screw, and bearing.
 
Event Description
It was reported that while a patient was sitting on the labor and delivery bed, the bolt that attaches the foot crank weldment on the patient left side came out, causing the bed to unexpectedly lower.No adverse consequence or clinically relevant delay in treatment was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LD304 MATERNITY BED
Type of Device
TABLE, OBSTETRICAL, AC-POWERED (AND ACCESSORIES)
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
mary klaver
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key7238806
MDR Text Key99002001
Report Number0001831750-2018-00071
Device Sequence Number1
Product Code HDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K950889
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number4701000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-