• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: A.R. MEDICOM INC. (SHANGHAI) CO.,LTD. HENRY SCHEIN; COTTON ROLLS NON-STERILE # 2 MEDIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

A.R. MEDICOM INC. (SHANGHAI) CO.,LTD. HENRY SCHEIN; COTTON ROLLS NON-STERILE # 2 MEDIUM Back to Search Results
Catalog Number 102-5455
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Skin Irritation (2076); Wheal(s) (2241)
Event Date 06/17/2015
Event Type  malfunction  
Manufacturer Narrative
On 14-jul-2015, a medwatch was reported to the fda by (b)(4), and refers to amd medicom sterile (b)(4) cotton rolls economic non sterile 102-5455 in which the following was reported: "two female patients had experienced blistering and redness immediately after a routine procedure.As a result both female patients were prescribed kenalog cream to help treat the blistering and redness.Patient 1: female, (b)(6), patient 2: female, age not disclosed" the (b)(4) complaint was received by a.R.Medicom (b)(4) on 23-jun-2015 and has been logged as product quality complaint (b)(4) with the following additional information being provided: "one staff member, a female in her 50's touched her lip and turned bright red.The doctor advised that other unknown amount of patients had similar reactions, not sure of how many/sex/ ages.No further treatment required.A staff person had to have kenalog administer to blisters as was the same with one other patient, was administered kenalog".A request for an investigation was immediately made to the manufacturing site (b)(4).At the time of reception of the quality complaint, an evaluation was performed by amd medicom and it was determined that additional information was required in order to complete the investigation.Therefore, a questionnaire was forwarded to the complainant on 29-jun-2015 and a response obtained on 08-jul-2015 as detailed below: - have you changed your ordering recently (different cotton roll type)? have always used this type of cotton roll.- what and how do you store your cotton rolls (in a case, box, drawer? typically store in the box they come in or a drawer.- can you describe more the symptoms? as the cotton roll touched the staff members lip, it turned red and also had blisters develop.- how many patients had this reaction? had quite a few patients with the same symptoms however only two needed prescription kenalog cream.Not specific about other person (age or sex).- sex of each patient: one staff member is a female, other patient did not specify.- was it the first time the patient used this lot of cotton rolls? yes, the first time for this lot.- was it the first time the patient used this type of cotton rolls? not the first time they used this type of cotton roll.- upon using the cotton rolls, after what duration of time did the symptoms appear (e.G.5, 10 minutes)? the symptoms were immediate about five minutes.- once the cotton roll was removed, how long did the symptoms persist? customer did not indicate how long after the symptoms persisted.- did the symptoms subside or disappear due to any treatment (medication?)? both a female staff member and a patient needed to get prescription kenalog cream.- was the patient hospitalized? no one needed to be hospitalized.- did the patient have a history of allergy? neither the staff member or patient have a history or allergy.Even though requests were made to the complainant for the return of the item code 102-5455 lot 18636, at that time, no product was returned by the complainant or samples provided.Therefore, retained samples received from the manufacturing site were sent to a third party contract laboratory 05-aug-2015 in order to perform a cytotoxicity test.On 14-aug-2015, the final report was received from the third party external contract laboratory which indicated that the cotton rolls retained samples pass the cytotoxicity test with a score of 0.Based on the review of the above responses obtained from the complainant, the cytotoxicity testing and after further investigation, a second set of questions were submitted for a response on 17-aug-2015: - was this lot stored in the original box or drawer? typically they store these in the box they come in, some are in drawers.- did they still have a stock of this specific lot? if yes, did they still use it? we have sent a second notification to our customer to return the affected product; however we have not yet received it as of today.- could you please describe their routine procedure? they were not specific in providing the routine procedure.Review of the dhr indicated that lot 18636 was manufactured in compliance with current procedures and internal specifications.No deviations were noted during the manufacturing process.It should be mentioned that as part of the manufacturing process, the cotton rolls were pretreated by gamma prior to shipment.A review of all complaints received by amd medicom over the last two (2) years has indicated that there was no other complaint received with respect to this particular product and the nature of the complaint.Based on the above investigation and the conclusive cytotoxicity data (that the cotton rolls from lot 18636 resulted in a score 0), it can be considered that the reaction is not linked to the product itself but most probably related to the manner the cotton rolls were stored and/or manipulated while treating the patients.Note: the medwatch should have completed and submitted to fda on aug, 2015.Due to miscommunication between (b)(4) and amd medicom, the final conclusion to the medwatch was not submitted as required.It was understood by amd medicom at that time that the conclusion of the medwatch was to be sent by (b)(4).Corrective actions have been implemented internally to avoid any future misunderstanding and prompt completion of any medwatch.
 
Event Description
"two female patients had experienced blistering and redness immediately after a routine procedure.As a result both female patients were prescribed kenalog cream to help treat the blistering and redness.Patient 1: female, (b)(6), patient 2: female, age not disclosed".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HENRY SCHEIN
Type of Device
COTTON ROLLS NON-STERILE # 2 MEDIUM
Manufacturer (Section D)
A.R. MEDICOM INC. (SHANGHAI) CO.,LTD.
290-302 xiangche road
songjang district,
shanghai, china 20161 1
CH  201611
Manufacturer (Section G)
A.R. MEDICOM INC. (SHANGHAI) CO.LTD.
290-302 xiangche road
sonjang district
shanghai, china
CH  
Manufacturer Contact
dominique abecassis
1200 55th avenue
lachine (montreal), quebec H8T 3-J8
CA   H8T 3J8
MDR Report Key7239338
MDR Text Key99330096
Report Number9611959-2015-00001
Device Sequence Number1
Product Code EFN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Dentist
Type of Report Initial
Report Date 07/14/2015,11/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Dentist
Device Catalogue Number102-5455
Device Lot Number18636
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2015
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA07/14/2015
Distributor Facility Aware Date06/23/2015
Event Location Other
Date Report to Manufacturer07/16/2015
Date Manufacturer Received06/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/25/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age50 YR
-
-