It was reported that during a benign prostatic hyperplasia (bph) procedure, the fiber broke after 24,000 joules and 4 minutes of use.The fiber was exchanged, and the procedure was completed utilizing the second fiber.There was no consequences for the patient.
|
Product evaluation: failure analysis for fiber (b)(4): the glass cap exhibits moderate devitrification at output window; the metal cap exhibits mild detritus adhesion on surface; the fiber is broken within the outer flow tubing at 2.25 inches from the control knob, between distal tip and control knob, and exhibits indication of bending, crushed, and melting at break location; the fiber was tested with hene laser fixture, aim beam is present at break location; the outer flow tubing open end exhibits minor scratch marks.Probable root cause: based on the device analysis, the probable root cause of the failure is: excessive bending.
|