• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) GREENLIGHT MOXY FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) GREENLIGHT MOXY FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0010-2400
Device Problems Bent (1059); Break (1069); Melted (1385)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/18/2018
Event Type  malfunction  
Event Description
It was reported that during a benign prostatic hyperplasia (bph) procedure, the fiber broke after 24,000 joules and 4 minutes of use.The fiber was exchanged, and the procedure was completed utilizing the second fiber.There was no consequences for the patient.
 
Manufacturer Narrative
Product evaluation: failure analysis for fiber (b)(4): the glass cap exhibits moderate devitrification at output window; the metal cap exhibits mild detritus adhesion on surface; the fiber is broken within the outer flow tubing at 2.25 inches from the control knob, between distal tip and control knob, and exhibits indication of bending, crushed, and melting at break location; the fiber was tested with hene laser fixture, aim beam is present at break location; the outer flow tubing open end exhibits minor scratch marks.Probable root cause: based on the device analysis, the probable root cause of the failure is: excessive bending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT MOXY FIBER OPTIC
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer Contact
alyson harris
3070 orchard drive
san jose, CA 95134
4084563300
MDR Report Key7243590
MDR Text Key99783992
Report Number2937094-2018-00129
Device Sequence Number1
Product Code GEX
UDI-Device Identifier00878953005515
UDI-Public00878953005515
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/09/2019
Device Model Number0010-2400
Device Catalogue Number0010-2400
Device Lot Number21355245
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-