• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH & LOMB INCORPORATED PEROXICLEAR HYDROGEN PEROXIDE(MILAN); ACCESSORIES, SOFT LENS PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH & LOMB INCORPORATED PEROXICLEAR HYDROGEN PEROXIDE(MILAN); ACCESSORIES, SOFT LENS PRODUCTS Back to Search Results
Device Problem Product Quality Problem (1506)
Patient Problems Red Eye(s) (2038); Burning Sensation (2146)
Event Type  malfunction  
Manufacturer Narrative
A practitioner reported a patient experienced redness and stinging after use of the product.The (b)(6) describes the scope of injury associated with hydrogen peroxide exposure to the ocular tissue, ¿hydrogen peroxide is irritating to the eyes with a burning sensation, conjunctival hyperemia, lacrimation and severe pain which resolves within a few hours.There are rare cases of temporary corneal injury resulting from the application of 3% solution to the eye on contact lenses including punctate staining of the cornea, decreased vision, corneal opacity and edema.¿ the symptoms reported could be associated with application of un-neutralized hydrogen peroxide coming in contact with the eye.
 
Event Description
A practitioner reported a patient experienced redness and stinging after use of the product.The patient was instructed to discard the product and to use a different company's solution.The practitioner reports the patient's symptoms disappeared after switching solution.The complaint suggests the device may have malfunctioned as a result of variability of neutralization.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEROXICLEAR HYDROGEN PEROXIDE(MILAN)
Type of Device
ACCESSORIES, SOFT LENS PRODUCTS
Manufacturer (Section D)
BAUSCH & LOMB INCORPORATED
1400 north goodman street
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB-IOM S.P.A
via pasubio 34
20050 macherio
milan 20846
IT   20846
Manufacturer Contact
tes proud
1400 north goodman street
rochester, NY 14609
5853388549
MDR Report Key7244572
MDR Text Key99789850
Report Number0001313525-2018-00034
Device Sequence Number1
Product Code LPN
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
K132216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial
Report Date 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/09/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberZ-0651-2017
Patient Sequence Number1
Patient Age29 YR
-
-