• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP* APPLIER W/MECH STOP; LIGACLIP APPLIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP* APPLIER W/MECH STOP; LIGACLIP APPLIER Back to Search Results
Catalog Number LX107
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Only event year known: 2018.
 
Event Description
It was reported that during an unknown procedure the clip applier was noted to apply clips unevenly.It was not reported how the procedure was completed.There were no patient consequences reported.
 
Manufacturer Narrative
Device analysis: the analysis results confirmed that the lx107 device was returned non functional as the jaws were misaligned; therefore, the clips could not be loaded into the jaws.No conclusion could be reached as to what may have caused the found condition of the jaws.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The batch history records were reviewed and certed by external manufacturing that the manufacturing criteria was met prior to the release of the equipment. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
LIGACLIP APPLIER
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7245341
MDR Text Key99308688
Report Number3005075853-2018-07801
Device Sequence Number1
Product Code GDO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberLX107
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received02/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-