• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: QUEST MEDICAL, INC. MPS 2 CONSOLE; CPBP HEAT EXHANGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

QUEST MEDICAL, INC. MPS 2 CONSOLE; CPBP HEAT EXHANGER Back to Search Results
Model Number 5201260
Device Problems Air Leak (1008); Device Displays Incorrect Message (2591); Device Sensing Problem (2917)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
A field service engineer from quest medical will travel to the facility to evaluate the device.A follow-up medwatch will be submitted if additional information becomes available.
 
Event Description
A report was received regarding an alleged issue encountered during use of the console.The report states that the console displayed the error code "is there air in heat exchanger" and the vent line had to be manually clamped to complete the case.There were no reported patient complications.
 
Manufacturer Narrative
The device was evaluated and the fluid level sensor was found to be defective.The sensor was replaced and the console was tested with no further errors.Raised edges on the sensor is suspected to be the root cause of this defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MPS 2 CONSOLE
Type of Device
CPBP HEAT EXHANGER
Manufacturer (Section D)
QUEST MEDICAL, INC.
one allentown parkway
allen TX 75002
Manufacturer (Section G)
QUEST MEDICAL, INC.
one allentown parkway
allen TX 75002
Manufacturer Contact
tosan onosode
one allentown parkway
allen, TX 75002
9723326338
MDR Report Key7246246
MDR Text Key99795123
Report Number1649914-2018-00013
Device Sequence Number1
Product Code DTR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K041979
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Model Number5201260
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/18/2018
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-