VYAIRE MEDICAL, INC 3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100B
Device Problem Device Stops Intermittently (1599)
Patient Problem Hypoxia (1918)
Event Date 01/10/2018
Event Type  Injury  
Manufacturer Narrative
Vyaire medical complaint (b)(4).In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
 
Event Description
The customer reported that the ventilator shut down, while the device was in use on a patient.The patient had hypoxmia and had to be resuscitated to maintain oxygen level.The patient was placed on another oscillator.When the customer evaluated the device after the reported issue, fluid was noticed in the pressure line inside the oscillator, above the pressure transducer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
kristin graf
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key7248267
MDR Text Key99235564
Report Number2021710-2018-07399
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100B
Device Catalogue Number770155
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/10/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-