Catalog Number SGC0302
Device Problems Mechanical Problem (1384); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/15/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event Description
This is filed to report the inability to straighten the steerable guide catheter tip.It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4.The steerable guide catheter (sgc) was prepared without issue per the instructions for use.The +/- knob was working fine.During groin insertion, the +/- knob was turned from 1/2 to 3/4 in the minus direction when a clicking noise was heard, the knob moved freely in the minus direction and the guide would not straightened anymore.The minus function was not working; however, the plus direction was still working.The sgc was exchanged for a new one and the procedure continued without further issue.1 clip was implanted reducing mr to grade <1.No adverse patient effects or a clinically significant delay in the procedure were reported.No additional information was provided.
 
Manufacturer Narrative
(b)(4).Evaluation summary: the device was returned and investigated.The reported mechanical issue was confirmed due to a broken cable, which resulted in the steerable guide catheter (sgc) unable to curve.The reported device operates differently than expected (noise) could not be replicated in a testing environment as it was likely a symptom of the cable break.A review of the lot history record revealed no manufacturing nonconformities reported to this lot.Additionally, a review of the complaint history identified no other incidents reported from this lot.All available information was investigation and the reported sgc unable to curve guide and noise appear to be related to the broken cables; however, a conclusive cause for the cable break could not be determined.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
STEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
MENLO PARK, CA REG# 3005070406
abbott vascular
26531 ynez rd.
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key7249413
MDR Text Key99798096
Report Number2024168-2018-00894
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/12/2018
Device Catalogue NumberSGC0302
Device Lot Number70511U206
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/26/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-