• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 90CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 90CM LENGTH; SCS LEAD Back to Search Results
Model Number 3189
Device Problem Kinked (1339)
Patient Problem Inadequate Pain Relief (2388)
Event Date 12/20/2017
Event Type  Injury  
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
Device 1 of 2.Reference mfr.Report: 1627487-2018-01022.It was reported the patient experienced ineffective stimulation, as the leads were auto reducing.Reprogramming attempts were unsuccessful.As a result, the patient underwent scs system replacement.During the surgery, the leads were found to be kinked under the site of the anchors on both leads.
 
Event Description
Device 1 of 2.Reference mfr.Report: 1627487-2018-01022.Follow up information identified postoperatively, effective therapy was restored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 90CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7252140
MDR Text Key99362858
Report Number1627487-2018-01021
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/02/2018
Device Model Number3189
Device Lot Number5586720
Other Device ID Number05415067017314
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/02/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 1192 (X2), SCS ANCHOR; MODEL 3662, SCS IPG
Patient Outcome(s) Other;
Patient Age37 YR
-
-