• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPTA CRTP; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPTA CRTP; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION Back to Search Results
Model Number W1TR01
Device Problems Electrical /Electronic Property Problem (1198); Pacing Problem (1439)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/05/2017
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: 5076-52 lead, implanted: (b)(6) 2007; 305c27 tissue valve, implanted: (b)(6) 2007; 383069 lead, implanted (b)(6) 2017.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that after the implant of the device, rapid right ventricular pacing was seen immediately following atrial anti-tachycardia pacing (atp).There were also long pauses following the atrial atp.Atrial atp was then turned off for the patient.The device remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Upon investigation it was determine that the device was acting within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEPTA CRTP
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7255032
MDR Text Key99533452
Report Number3004209178-2018-02578
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00643169735637
UDI-Public00643169735637
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/28/2018
Device Model NumberW1TR01
Device Catalogue NumberW1TR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/05/2018
Date Device Manufactured06/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age77 YR
-
-