The damaged section of the 70cc tah-t cannula was repaired by the hospital staff and remains on the patient; therefore no piece of cannula was returned to syncardia for evaluation.Syncardia conducted a review of the tah-t device history record (dhr) and sterilization work order and confirmed that all manufacturing was performed to specifications, and the 70cc tah-t met all specified requirements prior to shipment.Syncardia has a corrective and preventive action (capa) to address the cannula tear issue.The syncardia freedom driver system guidebook for patients and caregivers - us, instructs the patients and caregivers to inspect the drivelines and cannulae daily.If a hole is observed in the drivelines or cannulae, household tape must be applied to temporarily repair the hole and the hospital contact person must be contacted to arrange for a hospital-based evaluation.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
|
The customer, a syncardia certified hospital, reported a tear on the right side of the cannula.The customer also reported that the cannula was repaired by applying rescue tape around the cannula in the area of the tear.There was no reported adverse patient impact as a result of the cannula tear and subsequent repair.
|