SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T); BIVENTRICULAR REPLACEMENT DEVICE Back to Search Results
Model Number 500101-001
Device Problem Physical Property Issue (3008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/17/2018
Event Type  malfunction  
Manufacturer Narrative
The damaged section of the 70cc tah-t cannula was repaired by the hospital staff and remains on the patient; therefore no piece of cannula was returned to syncardia for evaluation.Syncardia conducted a review of the tah-t device history record (dhr) and sterilization work order and confirmed that all manufacturing was performed to specifications, and the 70cc tah-t met all specified requirements prior to shipment.Syncardia has a corrective and preventive action (capa) to address the cannula tear issue.The syncardia freedom driver system guidebook for patients and caregivers - us, instructs the patients and caregivers to inspect the drivelines and cannulae daily.If a hole is observed in the drivelines or cannulae, household tape must be applied to temporarily repair the hole and the hospital contact person must be contacted to arrange for a hospital-based evaluation.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported a tear on the right side of the cannula.The customer also reported that the cannula was repaired by applying rescue tape around the cannula in the area of the tear.There was no reported adverse patient impact as a result of the cannula tear and subsequent repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T)
Type of Device
BIVENTRICULAR REPLACEMENT DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key7260142
MDR Text Key99996938
Report Number3003761017-2018-00044
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003008
UDI-Public(01)00858000003008
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial
Report Date 01/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Model Number500101-001
Device Catalogue Number500101
Device Lot Number107138
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/17/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
-
-