SMITH & NEPHEW, INC. ULTRA FAST-FIX ASSEMBLY - CURVED; STAPLE, FIXATION, BONE Back to Search Results
Model Number 72201491
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/06/2018
Event Type  malfunction  
Event Description
It was reported that during the surgery, after t1 was stripped off successfully, t2 failed to be stripped off by manual advancement.A backup device was used to complete the surgery.No patient injury was reported.
 
Manufacturer Narrative
One 72201491 ultra-fast-fix needle delivery system device received.The complaint said: ¿it was reported that during the surgery, after t1 was stripped off successfully, t2 failed to be stripped off by manual advancement.¿ the blue split cannula was returned separated from the delivery needle.The white depth/penetration limiter was returned and has been trimmed.T1 is missing.T2 and partial suture are separated from the needle.The device is severely disfigured.The actuator¿s manual advancement button was received fully forward.The device is beyond mechanical testing due to manipulation and damages attained.Twist or bend can interfere with advancement and removal of t¿s.No root cause related to the manufacture of the device was established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA FAST-FIX ASSEMBLY - CURVED
Type of Device
STAPLE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7270507
MDR Text Key100145058
Report Number1219602-2018-00221
Device Sequence Number1
Product Code JDR
UDI-Device Identifier03596010597205
UDI-Public(01)03596010597205(17)220515(10)50664312
Combination Product (y/n)N
PMA/PMN Number
K972326
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 03/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/12/2022
Device Model Number72201491
Device Catalogue Number72201491
Device Lot Number50664312
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/20/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
-
-