This is a retro-spective complaint review- the endoskeleton tas bone screw 2300-5530 (v13) is jammed through the endoskeleton tas implant cage window, 2312-0112-n((b)(4)).The bone screw is unable to be separated from the implant with use of normal instrumentation.The bone screw contained damaged threading which likely resulted from the bone screw incorrectly being inserted through the implants cage window.The bone screw's hex has damage and appears to be stripped.This damage could have been caused by the screw being incorrectly forced into position or while attempting to back the screw out.The implant has damage at the bone screw insertion holes and the implants cage window, the implant has been deformed in these areas reasonably caused by a bone screw being forced into an incorrect position.Endoskeleton tas surgical technique provides instructions for the proper placement of bone screws with tas implants.X-rays should be used to confirm the desired trajectories of the bone screws.If surgeon used x-rays to confirm trajectory of the bone screw in this case, surgeon would have been able to identify the incorrect trajectory of the bone screw prior to driving the screw into the jammed position within the implant.A review of the dhr was performed, the review revealed there were no anomalies or non-conformances generated during the manufacture of these devices that would attribute to this complaint.The device was found to be within specification during final inspection.
|
During a scoli deformity correction the surgeon was doing an alif l4/l5 procedure, somehow drove the screw through the back of the implant and were not able to back the screw out.The surgeon attempted to back out the cage with the screw in place.During the process something vascular was nicked, the vascular surgeon clamped the vessel and was able to stop the bleeding.Surgeon was able to put in another cage and complete the procedure.The vascular surgeon closed the patient and the surgery was completed.
|