(b)(4).The sample was returned for evaluation.A visual exam was performed and it was observed that the cuff was deformed.The silicone cuff material of the device looks sagged and lost its resilience due to stress.When compared with a retained sample, the shape of the material of the returned sample looked different than the retained sample.The checked valve of the returned device was functional; it could be inflated and deflated.In addition, the red plug vent is free from blockage and air could be released.A device history record (dhr) review was performed and there were no issues found.Air pressure is a known cause that leads to cuff herniation/internal rupture of components of the lma mask.During high temperature, vacuum, and humid environment, and air/moisture left inside the cuff will expand significantly and it will damage the properties of the components to an irreparable condition.It is highly suspected that some residual air/moisture was left inside the device while handling/reprocessing that caused the failure.The cuff must be completely deflated prior to autoclaving.
|