MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, 5MM MULTI-FUNCTION PEEK HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, 5MM MULTI-FUNCTION PEEK HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250181106
Device Problem Device Packaging Compromised (2916)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/18/2018
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.(b)(4).
 
Event Description
It was reported that the insulation had been compromised.
 
Manufacturer Narrative
The product was returned for investigation and the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: insulation pulling back at distal tip and the post is broken the failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root causes could be user misuse, excessive force, normal wear, or improper sterilization methods.Gtin: (b)(4).
 
Event Description
It was reported that the insulation had been compromised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 5MM MULTI-FUNCTION PEEK HANDLE
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key7278485
MDR Text Key100515469
Report Number0002936485-2018-00144
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250181106
Device Lot Number1447989¿H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received01/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-