Model Number PM2272
Device Problem Difficult to Interrogate (1331)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/07/2018
Event Type  malfunction  
Event Description
During follow-up, intermittent telemetry was noted on the device during interrogation.The device was explanted and replaced during an unrelated procedure.The patient was in stable condition following the replacement.
 
Manufacturer Narrative
Device image showed that it had high usage on high power telemetry in the field.Customer complaint of device being stuck at ¿updating¿ is consistent with that occurred due to device being switched to 8k telemetry to preserve battery power after extensive high power usage.Analysis performed indicated that device exhibited normal device characteristics, device was able to perform interrogation after each test without any anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7279847
MDR Text Key100474893
Report Number2017865-2018-02772
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2018
Device Model NumberPM2272
Device Lot NumberP000043962
Other Device ID Number05414734509589
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-