Model Number 37800 |
Device Problems
Failure to Deliver Energy (1211); Ambient Noise Problem (2877)
|
Patient Problems
Nausea (1970); Therapeutic Effects, Unexpected (2099)
|
Event Date 02/14/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a consumer regarding a patient with an implantable neurostimulator (ins) for gastric stimulation.It was reported that the patient had surgery and the surgery was not related to the device/therapy; however, the patient had an ultrasound done during the surgery and since the surgery, the patient was nauseous so they wanted to check the device to make sure it was still working.No further complications were reported/anticipated.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a consumer.It was reported that the patient went to see the implanting physician and the device was off; it was turned back on and reprogrammed, so the device was not damaged.It was noted that the only way it could have been turned off was by the ultrasound that was used during the surgery.It was reported that after the surgery, the patient was very sick.The illness resolved by turning on and reprogramming, the patient was doing well now, and the device really helped with their nausea.No further complications were reported/anticipated.
|
|