• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESTIVA 7100; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESTIVA 7100; ANESTHESIA GAS MACHINE Back to Search Results
Device Problems Device Inoperable (1663); Failure of Device to Self-Test (2937)
Patient Problem No Patient Involvement (2645)
Event Date 01/23/2018
Event Type  malfunction  
Manufacturer Narrative
The distributor performed a checkout of the equipment and confirmed the reported complaint.The drive gas pressure sensor switch was replaced, and the unit was returned to service.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.The distributor performed a checkout of the equipment and confirmed the reported complaint.The drive gas pressure sensor switch was replaced, and the unit was returned to service.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported the unit alarmed during preuse checkout and lost the ability to mechanically ventilate.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESTIVA 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7284180
MDR Text Key100654360
Report Number2112667-2018-00372
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K000706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 02/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/23/2018
Date Device Manufactured03/07/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-