• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. ETHILON SUTURE 18" (45CM) 4-0 BLK; SUTURE, NONABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. ETHILON SUTURE 18" (45CM) 4-0 BLK; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number 1692G
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/29/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported that the patient underwent an unknown dental surgery on (b)(6) 2018 and suture was used.During the procedure, the needle was easily detached from the suture and fell into the oral cavity when trying to use the needle-suture.It was not a control release needle.There were no adverse consequences to the patient.No additional information is available.
 
Manufacturer Narrative
(b)(4).Date sent to the fda: (b)(4) 2018.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.
 
Manufacturer Narrative
(b)(4).Actual product returned.No defects along of the strand were found; however, the end of the suture is severely damaged (flat) appears to be by the use of a surgical instrument this caused the needle was detached from the suture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHILON SUTURE 18" (45CM) 4-0 BLK
Type of Device
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
Manufacturer (Section G)
ETHICON INC.
road 183, km. 8.3
san lorenzo PR 00754
Manufacturer Contact
darlene kyle
p.o. box 151, route 22 west
somerville, NJ 08876-0151
9082182792
MDR Report Key7285590
MDR Text Key100778178
Report Number2210968-2018-70949
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K946173
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2021
Device Catalogue Number1692G
Device Lot NumberKLQ233
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received03/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-