Catalog Number 72200873
Device Problem Melted (1385)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported that sectors of the mcb have melted, it has been overloaded.No case involved.No patient injury was reported.
 
Manufacturer Narrative
A visual inspection was performed on the product and no issue was observed.There was no relationship found between the returned device and the reported incident.Product failed to power up.Cause of power up failure is a dead power supply.A known good power supply was installed and unit powered up.Then the unit failed for ¿footswitch required¿ error message in port a only.When an mdu was inserted into port a the motor could not be activated.Cause of malfunction in port a is a defective electronic component on the main pcb part number.Also a resistor was burnt.Reason for electronic component failure is unknown but a shorted out mdu is a possibility.Product passed functional testing with a known good power supply and main pcb installed.The complaint investigation has concluded the cause of the failure to be a defective electronic component.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DII CONTROLLER
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7286811
MDR Text Key100733121
Report Number1643264-2018-00156
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
PMA/PMN Number
K062849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200873
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received03/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-