• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801046
Device Problems Corroded (1131); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/29/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The field service representative (fsr) sent the user facility a loaner pump.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, an "overspeed" error message was displayed.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Evaluation is in progress, but not yet concluded.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) pressed the start button, the speed control knob increased and immediately an error message of "stopped: overspeed" was displayed.This is atypical as the pump should begin rotation.Corrosion of the motor controller pins in the control module was noted and likely the cause of the reported complaint.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7290100
MDR Text Key100908915
Report Number1828100-2018-00090
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K153376
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 05/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801046
Device Catalogue Number801046
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2018
Was the Report Sent to FDA? No
Date Manufacturer Received05/08/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-