Catalog Number PMX110 |
Device Problems
Aspiration Issue (2883); Device Operates Differently Than Expected (2913); No Pressure (2994)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/26/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
|
Event Description
|
During preparation for a thrombectomy procedure, the penumbra system aspiration pump max 110v (pump max) was powered on with the on/off button illuminated green; however, the pump max was unable to produce any vacuum.Therefore, the physician decided to only do a diagnostic and postponed the procedure.
|
|
Manufacturer Narrative
|
Results: the pump max was unable to be opened conventionally.The pump max was opened forcefully damaging some of the screw anchors.The interior of the device revealed unidentified black and yellow liquid residues throughout the pump housing and device components.The capacitor was found unscrewed and free from its housing.Yellow residue was found on the screw mounting holes and exiting the cooling fan air intake.Conclusions: evaluation of the returned pump max confirmed the device was unable to produce aspiration.The pump max housing was forcefully opened and liquid residues were found throughout the device.The capacitor was covered in a black grease-like substance and loose from its housing.If the capacitor was compromised, it would likely contribute to the inability of the pump assembly to power on and produce aspiration.The returned condition discovered inside the pump housing suggests the device was likely stored and cared for poorly.It is possible the device was opened and tampered with.Penumbra pumps are visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Search Alerts/Recalls
|