This report is being filed to provide additional information.Investigation: a review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.The run data file (rdf) was analyzed for this event.Signals in the rdf confirmed that the system operated as intended by detecting air in the return line.Review of the rdf indicated that during rinseback, approximately 100 minutes, the return line air detector started to detect small amounts of air in the return line.This continued to accumulate for the next 5 minutes, until the alarm was triggered during rinseback.The "air was detected in return line" alarm occurred at 105 minutes.Investigation is still in process.A follow-up report will be provided.
|
This report is being filed to provide additional information.Investigation: the customer stated that the operator noticed air bubbles in the return line past the rlad.The run data file (rdf) was analyzed to determine the alarms and signals from the return line air detector (rlad) in the optia.The rlad is situated at the exit of the return pump and is the last point of air detection in the optia circuit.Signals in the rdf showed an rlad air alarm and the operator was not able to complete rinseback because of the rlad air alarm.The patient showed no symptoms or signs attributable to air embolism.Per terumo bct's internal risk documentation, outgassing of cold replacement fluids can occur when the fluids are warmed through the blood warmer in tpe procedures.When blood warmer tubing is connected to the return line, any air bubbles could be infused into the patient.Based on the calculation of the maximum rate of air generated, a low level health hazard risk from outgassing in the reported case is expected.The absence of reported clinical symptoms when outgassing has been noted and supports this contention.Root cause: during tpe procedures on spectra optia, the replacement saline, albumin or plasma may be cold.If they are not allowed to warm to room temperature, and if the return blood is warmed, air bubbles may form.The reason for this "outgassing" is that gasses are more soluble in liquids at low temperatures than at higher temperatures.If at a low storage temperature air is available to dissolve in a fluid (albumin, plasma, etc.) and approaches its equilibrium solubility at that temperature, when the fluid is warmed the air will come out of solution, because its solubility is exceeded at the higher temperature.It is usually described as chains of very small bubbles or foam, which tend to rise toward the top of the tubing.The small bubbles may coalesce to form larger bubbles.Because cold ffp was used as part of the replacement fluid in the tpe procedure, the operators used an astotherm blood warmer in the return line in each of the procedures.The astotherm blood warmer consists of a length of tubing wrapped around a heated mandrel.Heat is transferred to the return-line blood to prevent the sensation of cool return blood.The root cause for the reported air in return line was due to outgassing of small bubbles out of solution due to cold replacement fluid being warmed.The system operated as intended.
|