The customer contacted the siemens customer care center (ccc) for the discordant phosphorus (phos) results obtained on the dimension vista instrument.Siemens healthcare diagnostics headquarters support center (hsc) concluded their investigation of the discordant phosphorus (phos) results.Headquarters support center (hsc) has reviewed the information provided as well as a review of instrument data.Based on review of calibration history, the customer calibrated phos using reagent lot 17291ac and used this calibration until (b)(6) 2018 when qc recovery was out of range high.The qc was run using flex sequence 104, well set 6;10;2.The issue was resolved by recalibrating phos using a new flex sequence 103 and well set 5;9;1.The discrepant patients were repeated and corrected reports were issued.After recalibration, the instrument is currently working as specified.No further action will be taken by hsc.No further evaluation of the device is required.
|