• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION ACTIVA SC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION ACTIVA SC Back to Search Results
Model Number NEU_INS_STIMULATOR
Device Problems Break (1069); High impedance (1291); Migration or Expulsion of Device (1395); Low Battery (2584); Material Deformation (2976)
Patient Problems Twiddlers Syndrome (2114); Therapeutic Response, Decreased (2271)
Event Date 11/25/2017
Event Type  Injury  
Manufacturer Narrative
Date of event.Please note that this date is based off of the date of publication of the article as the event dates were not provided in the published literature.Information references the main component of one of the systems involved in the reported events; other applicable components are: product id: neu_unknown_ext, lot# unknown, product type: extension.Product id: neu_unknown_lead, lot# unknown, product type: lead.Product id: 37612, lot# unknown, product type: implantable neurostimulator franzini, a., ranieri, r., gambini, o., messina, g.Manipulating an internal pulse generator until twiddler's syndrome in a patient treated with deep brain stimulation for obsessive-compulsive disorder.Acta neurochir (wien).2017.Doi: 10.1007/s00701-017-3412-9.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Summary: background: twiddler¿s syndrome consists of rotation or manipulation of an implantable pulse generator (ipg) in its subcutaneous pocket by a patient, thus causing hardware malfunction.Methods: this syndrome is being reported more frequently in patients treated with deep brain stimulation (dbs).Results: we report the case of a woman who had received bed nucleus of stria terminalis (bnst) electrodes for obsessive-compulsive disorder (ocd) and developed twiddler¿s syndrome a few months after surgery, causing hardware malfunction due to obsessive manipulation of the ipg.Conclusion: the patient did not have compulsions related to touching objects at admission, thus making it difficult to foresee and prevent ts.Reported events: a (b)(6)-year-old woman with bilateral deep brain stimulation (dbs) of the bed nucleus of stria terminalis (bnst) electrodes for obsessive-compulsive disorder (ocd) experienced depletion of their left implantable neurostimulator (ins) along with a worsening of their clinical presentation after approximately 6-7 months in which they had achieved ¿slow but evident clinical improvement in ocd and general functioning.¿ with the right ins also reportedly getting low the authors decided to replace them with two rechargeable ins¿s.They wrote that during the replacement surgery the left extension was observed to be extensively coiled on itself.The patient¿s family reported that the patient had frequently manipulated the device at both the thoracic and extension/lead connector positions, and the authors hypothesized that the braiding of the extension was caused by rotation of the ins in the loose connective tissue of the infraclavicular pocket.Postoperative impedance readings on the right side were normal, but high impedances were found on all contacts of the left lead.X-ray imaging confirmed there was braiding and fraying of the extension which continued to the extracranial part of the lead.Another surgery was performed to replace the left extension and to secure the ins to the muscle fascia with silk sutures.Six months later there had reportedly been no recurrence of twiddler¿s syndrome.The authors reported that the patient was implanted with an activa sc neurostimulator, which was later replaced with an activa rc rechargeable neurostimulator, but the model number of activa sc was not provided.It was not possible to ascertain any further specific device information from the article or to match the reported event with any previously reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA SC
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7298068
MDR Text Key101000482
Report Number3007566237-2018-00570
Device Sequence Number1
Product Code MFR
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
H050003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Reporter Occupation Physician
Type of Report Initial
Report Date 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNEU_INS_STIMULATOR
Device Catalogue NumberNEU_INS_STIMULATOR
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age51 YR
-
-