Results: the penumbra system ace 68 reperfusion catheter (ace68) was kinked approximately 35.0 and 76.0 cm from the hub.Conclusions: evaluation of the returned device revealed it was kinked.This damage may have occurred due to forceful manipulation of the ace68 at extreme angles during removal from the packaging.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|