• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE CLEARCUT KNIVES; KNIFE, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE CLEARCUT KNIVES; KNIFE, OPHTHALMIC Back to Search Results
Catalog Number 8065993048
Device Problems Bent (1059); Failure to Cut (2587)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/09/2018
Event Type  malfunction  
Manufacturer Narrative
A sample has been received by manufacturing that has not yet been evaluated.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A nurse reported that a knife did not cut well and the blade edge was bent upwards while making the incision during cataract surgery.An alternate knife was obtained in order successfully complete the procedure.There was no impact to the patient.
 
Manufacturer Narrative
One opened knife sample was received in a blister package with the lid stock taped to the blister for the report of the knife edge is bent and will not cut.The returned sample was visually inspected and was found to be nonconforming with a damaged tip and damaged cutting edges.Penetration testing could not be performed due to the damaged condition of the sample.A review of the device history record related to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A complaint history examination indicates that there are no additional complaints associated with the lot for the reported issue.The exact root cause could not be determined from the investigation performed.The damage to the returned sample is consistent with damage that can occur when the blade contacts a hard surface such as the protective blade tray when the product is improperly removed or inserted after use, from improper handling or from contact with another instrument during surgery or set-up.The exact root cause for the damaged knife sample is unknown therefore, specific action with regards to this complaint cannot be taken.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Any nonconformance, such as the damaged tip and damaged cutting edges exhibited on the returned opened sample, is removed from the lot and scrapped.Functional penetration testing is performed and monitored during the finishing process to ensure the sharpness of the product.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARCUT KNIVES
Type of Device
KNIFE, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LTD. - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7300551
MDR Text Key101254453
Report Number2523835-2018-00090
Device Sequence Number1
Product Code HNN
Combination Product (y/n)N
Reporter Country CodeCY
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Catalogue Number8065993048
Device Lot Number184224M
Other Device ID Number380659930485
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age66 YR
-
-