The percutaneous lead was tested for electrical continuity in the condition that it was received and did not reveal any discontinuities or shorts.Examination of the percutaneous lead noted several tears under the rescue tapes.However, the underlying bionate was free of damage.Upon removal of the silicone sleeves, overall metal braided shield deterioration was noted.In addition, a minor kink was observed approximately 15 inches from the metal connector.The silicone jacket, clear bionate, and metal braided shield were removed in order to examine the underlying wires.Visual inspection of the percutaneous lead did not reveal any evidence of mechanical damage or breakdown of the wire insulation.The percutaneous lead was submerged in a saline bath for high-potential testing to check for current leakage through each wire's insulation.The test did not reveal any insulation breaches that would have contributed to an electrical short.The patient was placed on regular grounded patient cable post repair and discharged.The patient remains ongoing on the left ventricular assist system (lvas) and no further related issues have been reported.A review of the device history records revealed no deviations from manufacturing or quality assurance specifications.No further information was provided.The manufacturer is closing the file on this event.
|