• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754LWWL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754LWWL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-754LWWL
Device Problems Break (1069); Positioning Problem (3009)
Patient Problem Hyperglycemia (1905)
Event Date 02/10/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
It was reported that the reservoir ring was damaged and small piece was missing.The customer¿s blood glucose level was 300 mg/dl.The customer also reported that the reservoir was able to lock but barely in place when inserted in insulin pump.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Pump had cracked battery tube threads, reservoir tube lip almost completely broken off and test reservoir unable to lock into place, broken belt clip slot, minor scratched display window and stained address/serial number label.Pump passed the displacement test, prime and excessive no delivery test.Unable to perform the rewind, basic occlusion and occlusion test due to reservoir tube lip anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754LWWL
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7302835
MDR Text Key101156741
Report Number3004209178-2018-53628
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169515727
UDI-Public(01)00643169515727
Combination Product (y/n)N
Reporter Country CodeSP
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-754LWWL
Device Catalogue NumberMMT-754LWWL
Device Lot NumberA2754LWWLJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-