• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEVRO CORP. SENZA; NEVRO SENZA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEVRO CORP. SENZA; NEVRO SENZA Back to Search Results
Model Number ACCK1015
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/31/2018
Event Type  malfunction  
Manufacturer Narrative
The insertion needle was not returned for analysis.This issue has been addressed with the manufacturer and corrective action has been implemented.Review of the severity and occurrence rate for this type of event shows that the overall risk is within acceptable limits.
 
Event Description
It was reported to nevro that the insertion needle detached from the hub during a trial procedure.It occurred when the physician was removing the needle following lead placement.The physician was able to remove the insertion needle without incident.The trial was completed with no further report of complication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENZA
Type of Device
NEVRO SENZA
Manufacturer (Section D)
NEVRO CORP.
1800 bridge parkway
redwood city CA 94065
Manufacturer Contact
catherine aronson
1800 bridge parkway
redwood city, CA 94065
6504333218
MDR Report Key7305355
MDR Text Key101350909
Report Number3008514029-2018-00129
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00813426020183
UDI-Public00813426020183
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date12/14/2018
Device Model NumberACCK1015
Device Catalogue NumberACCK1015
Device Lot Number9436656
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/14/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age54 YR
-
-